ABX Pentra 400 (2016-01-18) Jan 18, 2016 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type III
Lot or serial number Companies Brand name Model or catalog number Not Applicable Manufacturer Horiba ABX, Parc Euromedecine rue Du Caducee - BP 7290, Cedex 4, Montpellier 34184 FRANCE D. LACTIC ACID 260ML A11A01721 Not Applicable Manufacturer Horiba ABX, Parc Euromedecine rue Du Caducee - BP 7290, Cedex 4, Montpellier 34184 FRANCE B. ABX PENTRA 400 CLINICAL CHEMISTRY SYSTEM A11A01640 Not Applicable Manufacturer Horiba ABX, Parc Euromedecine rue Du Caducee - BP 7290, Cedex 4, Montpellier 34184 FRANCE C. uric acid cp 220T A11A01670 Not Applicable Manufacturer Horiba ABX, Parc Euromedecine rue Du Caducee - BP 7290, Cedex 4, Montpellier 34184 FRANCE A. CHOLESTEROL CP 380T A11A01634
Affected Products
CHOLESTEROL CP 380T
ABX PENTRA 400 CLINICAL CHEMISTRY SYSTEM
uric acid cp 220T
LACTIC ACID 260ML
N-acetylcysteine present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoantipyrine and phenol, and can produce falsely low results with the reagents using the Trinder reaction method.