SYNGO DYNAMICS VA10 VERSION (2016-02-01)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Siemens Healthcare GMBH
HENKESTRASSE 127
ERLANGEN
91052
GERMANY
SYNGO DYNAMICS VA10 VERSIONVA10

Introduction

Affected Products

SYNGO DYNAMICS VA10 VERSION

Reason

Siemens is notifying syngo Dynamics customers of a potential issue when using the measurement package of the VA10 version. If two measurements with the same name exist, instances of measurement will populate both measurements definitions. One of the definitions can be edited while one cannot be seen or edited. Both set of measurements are used in any associated calculations. Therefore if you edit or delete one of the measurement instances the one that is hidden will not receive the edits and all instances, edited and not edited, will be averaged together.