SYNGO IMAGING (2016-02-01)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
1094
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
SYNGO IMAGING10 014 063

Introduction

Affected Products

SYNGO IMAGING

Reason

Siemens is notifying customers using syngo Imaging Studio / advanced Studio version VB36C and older of the possibility of incorrect values for distance measurement. For measurement calculation, syngo Imaging VB36C and older version use the DICOM attribute "imager pixel spacing" (Tag (0018, 1164)) instead of "pixel spacing" (Tag (0028, 0030)). This may result in incorrect values for distance measurements. This happens if both DICOM attributes "imager pixel spacing" and "pixel spacing" are set and have different values.