Health Products

Trumpf Medizin Surgical Lights (2016-01-28)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
All LotsD. ILED 5 SURGICAL LIGHT
Manufacturer
Trumpf Medizin Systems GmbH T Co. Kg,
Karl-Zeiss-Str. 7-9,
Saalfeld,
07318
GERMANY
1500
All LotsE. TRUVIDIA
Manufacturer
Trumpf Medizin Systems GmbH T Co. Kg,
Karl-Zeiss-Str. 7-9,
Saalfeld,
07318
GERMANY
All LotsB. TRULIGHT 5000 SURGICAL LIGHT
Manufacturer
Trumpf Medizin Systems GmbH T Co. Kg,
Karl-Zeiss-Str. 7-9,
Saalfeld,
07318
GERMANY
T131
All LotsC. ILED 3 SURGICAL LIGHT
Manufacturer
Trumpf Medizin Systems GmbH T Co. Kg,
Karl-Zeiss-Str. 7-9,
Saalfeld,
07318
GERMANY
1300
All LotsTRULIGHT 3000 SURGICAL LIGHT
Manufacturer
Trumpf Medizin Systems GmbH T Co. Kg,
Karl-Zeiss-Str. 7-9,
Saalfeld,
07318
GERMANY
TE300H

Introduction

Affected Products
  1. TRULIGHT 3000 SURGICAL LIGHT
  2. TRULIGHT 5000 SURGICAL LIGHT
  3. ILED 3 SURGICAL LIGHT
  4. ILED 5 SURGICAL LIGHT
  5. TRUVIDIA

Reason

Manufacturer has received information stating the iLED and TruLight lighting systems have fallen from the central axis. The reported incidents occurred while the lights were being positioned for surgical procedures and have been involved in one injury to a user. The investigation of this issue has determined the root cause is due to the improper installation of the snap ring that can occur when a spring arm is attached to the central axis. The snap ring is located inside the interface and holds the spring arm in place. If the snap ring is installed improperly, the spring arm of the lighting system can descend over time, and eventually fall from the central axis. When the snap ring has been installed properly, the potential for the lighting system to fall does not exist.