MITRACLIP CLIP DELIVERY SYSTEM (2016-02-08)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturer.MITRACLIP CLIP DELIVERY SYSTEM
Manufacturer
Evalve, Inc. DBA Abbott Vascular Inc.
4045 CAMPBELL AVENUE
MENLO PARK
94025
California
UNITED STATES
CDS0201

Introduction

Affected Products

MITRACLIP CLIP DELIVERY SYSTEM

Reason

Abbott Vascular (AV) has recently received reports where the user attempted implanting a MitraClip, but the clip could not be detached from the delivery system due to a mandrel fracture. A mandrel fracture may occur if tension is present on the mandrel when turning the actuator knob to deploy the clip. Tension is present if the arm positioner is on the "closed" side of neutral, as opposed to being in the neutral position during clip deployment. The Field Safety Notice (FSN) is to alert users and provide an alternative deployment sequence that will ensure tension is not present in the system.