Abbott Vascular (AV) has recently received reports where the user attempted implanting a MitraClip, but the clip could not be detached from the delivery system due to a mandrel fracture. A mandrel fracture may occur if tension is present on the mandrel when turning the actuator knob to deploy the clip. Tension is present if the arm positioner is on the "closed" side of neutral, as opposed to being in the neutral position during clip deployment. The Field Safety Notice (FSN) is to alert users and provide an alternative deployment sequence that will ensure tension is not present in the system.