Health Products

DEFIBRILLATION PADS (2016-01-27)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
141125-02
Manufacturer
Cardiac Science Corporation
N7W22025 JOHNSON DRIVE
WAUKESHA
53186
Wisconsin
UNITED STATES
DEFIBRILLATION PADS9131

Introduction

Affected Products

DEFIBRILLATION PADS

Reason

Cardiac Science has determined that the affected lot may exhibit higher than expected electrical impedance over time.  If electrode impedance becomes too high, the G3 AED will fail the electrode self-test performed by the AED and it will not be rescue ready.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime