Affected Products
A. PHYSIO-CONTROL QUIK-COMBO PACING/DEFIBRILLATION/ECG ELECTRODES WITH EDGE SYSTEM TECHNOLOGY AND REDI-PAK PRECONNECT SYSTEM
Physio-Control's supplier of QUIK-COMBO adult, pacing/defibrillation/ECG Electrodes with EDGE System technology and REDI-PAK preconnect system is voluntarily recalling specific production lots. The defibrillation electrodes are used in conjunction with certain Physio-Control LIFEPAK products. This voluntary recall is being conducted due to a low-level potential for damage to the wire insulation during the manufacturing process of the affected lots. The use of products with this condition may result in a potentially increased risk for reduced or no patient therapy, arcing of current, sparking, and patient and/or clinician burns.