Introduction

Affected Products

VS200 RADIOGRAPHIC WALL STAND

Reason

Del Medical has determined that in rare cases, a fastener within the internal counterbalance system may fail. In the event of a fastener failure, the wallstand image receptor will descend to the lowest position and, cannot be used for x-ray examination until repairs are completed.