ALLEGRA X-5 (2015-02-08)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
322/13
Manufacturer
Beckman Coulter
Inc.
250 S. Kraemer Blvd.
92821
UNITED STATES
A. ALLEGRA X-5B30593

Introduction

Affected Products

A. ALLEGRA X-5

Reason

A defective rotor yoke allows improper rotor seating on the hub resulting in possible rotor failure if the buckets are loaded in an imbalanced configuration. If the rotor becomes unbalanced, it may vibrate excessively and/or exhibit smoke if the rotor is rubbing inside the centrifuge.