Health Products

AISYS CS2 ANESTHESIA MACHINE MAIN UNIT (2016-02-18)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: Healthcare Professionals, General Public, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All Lots
Manufacturer
Datex-Ohmeda
3030 Ohmeda Drive, P.O. Box 7550,
Madison
53707-7550
Wisconsin
UNITED STATES
A. AISYS CS2 ANESTHESIA MACHINE MAIN UNIT1011-9050-000

Introduction

Affected Products

A. AISYS CS2 ANESTHESIA MACHINE MAIN UNIT

Reason

GE Healthcare has recently become aware of a potential safety issue involving a momentary, self-correcting anesthetic agent bolus when using 21% oxygen on all Aisys CS2 and upgraded Aisys anesthesia devices. Delivery of this momentary bolus of anesthetic agent is potentially hazardous because it could lead to hypotension in certain vulnerable pediatric patients when 21% oxygen (air only) is used. There have been no injuries reported as a result of this issue.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime