Siemens Healthcare Diagnostics has confirmed a lot to lot bias with calibrator E kit for the ADVIA Centaur Testosterone assay.
Internal investigation demonstrates:
an average positive bias of 17% across the assay range with calibrator E kit lots ending in 42 as compared to values generated with the master curve.
an average negative bias of 7% across the assay range with calibrator E kit lots ending in 43 as compared to values generated with the master curve.
the positive bias of calibrator E kit lots ending in 42 to the master curve compared to the negative bias of calibrator E kit lots ending in 43 to the master curve is the main driver for the negative bias between values generated with calibrator E kit lots ending in 42 and calibrator E kit lots ending in 43.
The bias of the calibrators to the master curve has been corrected with calibrator E kit lots ending in 44.
The performance of the testosterone assay on the ADVIA Centaur CP is not affected.
Calibrator E is also used for calibration of the ADVIA centaur systems cortisol and progesterone assays, however the performance of these assays is not affected.