ADVIA Centaur System Calibrator (2016-02-16)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES
B. ADVIA CENTAUR SYSTEM-CALIBRATORS04634452
More than 10 numbers, contact manufacturer.
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES
A. ADVIA CENTAUR CP SYSTEM-CALIBRATOR04634452

Introduction

Affected Products
  • A. ADVIA CENTAUR CP SYSTEM-CALIBRATOR
  • B. ADVIA CENTAUR SYSTEM-CALIBRATORS

Reason

Siemens Healthcare Diagnostics has confirmed a lot to lot bias with calibrator E kit for the ADVIA Centaur Testosterone assay.
Internal investigation demonstrates: 
  • an average positive bias of 17% across the assay range with calibrator E kit lots ending in 42 as compared to values generated with the master curve. 
  • an average negative bias of 7% across the assay range with calibrator E kit lots ending in 43 as compared to values generated with the master curve. 
  • the positive bias of calibrator E kit lots ending in 42 to the master curve compared to the negative bias of calibrator E kit lots ending in 43 to the master curve is the main driver for the negative bias between values generated with calibrator E kit lots ending in 42 and calibrator E kit lots ending in 43.


The bias of the calibrators to the master curve has been corrected with calibrator E kit lots ending in 44.

The performance of the testosterone assay on the ADVIA Centaur CP is not affected.

Calibrator E is also used for calibration of the ADVIA centaur systems cortisol and progesterone assays, however the performance of these assays is not affected.