VITROS Immunodiagnostic Products Total beta-hCG II (Reagent Pack and Calibrator)
6802220
Introduction
Affected products
VITROS Immunodiagnostic Products Total beta-hCG II (Reagent Pack and Calibrator)
Reason
Ortho Clinical Diagnostics (Ortho) has confirmed that when testing was conducted using lot numbers 1410, 1440, 1460 and 1470, VITROS systems generated results within the measuring range for samples known to not contain measurable hCG. Customers reported that their VITROS system reported results up to approximately 7.40 mIU/mL (IU/L) for patient samples that should have been less than the measuring range of the assay (<2.39 mIU/mL (IU/L)).