VITROS Immunodiagnostic Products Total beta-hCG II (Reagent Pack and Calibrator) (2016-02-18)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: Healthcare Professionals, General Public, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
1410
Manufacturer
ORTHO-CLINICAL DIAGNOSTICS
FELINDRE MEADOWS, PENCOED, BRIDGEND
VITROS Immunodiagnostic Products Total beta-hCG II (Reagent Pack and Calibrator)6802220

Introduction

Affected products
  1. VITROS Immunodiagnostic Products Total beta-hCG II (Reagent Pack and Calibrator)

Reason

Ortho Clinical Diagnostics (Ortho) has confirmed that when testing was conducted using lot numbers 1410, 1440, 1460 and 1470, VITROS systems generated results within the measuring range for samples known to not contain measurable hCG. Customers reported that their VITROS system reported results up to approximately 7.40 mIU/mL (IU/L) for patient samples that should have been less than the measuring range of the assay (<2.39 mIU/mL (IU/L)).