Health Products

CIOS ALPHA (2016-02-23)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
11137
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
CIOS ALPHA10308191

Introduction

Affected Products
  1. CIOS ALPHA

Reason

Siemens is notifying customers of the following potential software issues with the Cios Alpha:
  1. If the system language is set to English and the yyyymmdd date format is used, the patient's date of birth received through the DICOM worklist and query/retrieve will be displayed with month and day reversed for patients who have birthdays on days 1 through 12. Birth dates for days 13 through 31 are displayed correctly. The issue only appears to happen for days 1 through 12.
  2. System crash may occur during DVD live recording if radiation is released while the DVD icon on the FLC status bar is blinking. In this case error messages will be displayed and a system reset is necessary.
  3. Sporadically the system will not be ready for radiation during reconnection of the trolley to the main unit. A system reboot is necessary.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime