Introduction

Affected Products
  1. RAYSTATION

Reason

Issue found with some of the tools used for drawing in and interacting with the patient views and beam's eye views. The issue concerns all tools that are used with the left mouse button held down. These tools can become unsynchronized with the stored data if other commands such as for example a right mouse click are used at the same time as the left mouse button is held down.