Health Products

Instrument Manager (IM) (2016-01-29)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not applicable
Manufacturer
Data Innovations LLC
120 Kimball Avenue Suite 100
South Burlington
Vermont
UNITED STATES
Instrument Manager (IM)8.12.10

Introduction

Affected Products

Instrument Manager (IM)

Reason

Importing error code mapping into Instrument Manager (IM) v 8.12.10 and higher may clear another configuration's error code mapping. This would cause results that should be flagged from that configuration to have no error code on results, and it could have an effect on rules that evaluate the error code(s).
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime