SIGNA PIONEER (2016-03-03)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
001719TA4
Manufacturer
GE Healthcare Japan Corporation
7-127 Asahigaoka, 4 Chome, Hinoshi
Tokyo
191-8503
JAPAN
A. SIGNA PIONEERSIGNA PIONEER

Introduction

Affected Products     

A. SIGNA PIONEER

Reason

GE Healthcare has recently become aware of a potential safety issue with a missing ferrous warning label on SIGNA Pioneer table control box (TCB). Ferrous material in the TCB could cause unexpected attraction to the MR system's high-strength magnetic field during servicing activities. This could create a potential risk of serious injury to service personnel who attempt to inappropriately remove the TCB from the magnetic field. There have been no events reported as a result of this issue. Clinical scanning of patients is not affected by this safety notification.