ALLURA XPER SYSTEMS (2016-03-10) Mar 10, 2016 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : Healthcare Professionals, General Public, Hospitals
Department : Health Canada
Hazard classification : Type III
Lot or serial number Companies Brand name Model or catalog number 146 Manufacturer Philips Medical Systems Nederland B.V. Veenpluis 4-6 Best 5684 NETHERLANDS F. ALLURA XPER FD20 OR Table 454 More than 10 numbers, contact manufacturer. Manufacturer Philips Medical Systems Nederland B.V. Veenpluis 4-6 Best 5684 NETHERLANDS D. ALLURA XPER FD10 ALLURA XPER FD10 More than 10 numbers, contact manufacturer. Manufacturer Philips Medical Systems Nederland B.V. Veenpluis 4-6 Best 5684 NETHERLANDS E. ALLURA XPER FD20 SYSTEM - Main Unit ALLURA XPER FD20 168 Manufacturer Philips Medical Systems Nederland B.V. Veenpluis 4-6 Best 5684 NETHERLANDS B. ALLURA XPER FD20/10 SYSTEM - Main Unit ALLURA XPER FD20/10 171 Manufacturer Philips Medical Systems Nederland B.V. Veenpluis 4-6 Best 5684 NETHERLANDS C. ALLURA XPER FD20/20 ALLURA XPER FD20/20 154 Manufacturer Philips Medical Systems Nederland B.V. Veenpluis 4-6 Best 5684 NETHERLANDS A. ALLURA XPER FD10/10 - Main Unit ALLURA XPER FD10/10
Affected products
ALLURA XPER FD10/10 - Main Unit
ALLURA XPER FD20/10 SYSTEM - Main Unit
ALLURA XPER FD20/20
ALLURA XPER FD10
ALLURA XPER FD20 SYSTEM - Main Unit
ALLURA XPER FD20 Operating Room Table
Philips Healthcare has discovered through customer complaints and internal testing an intermittent electronic product defect in accordance with Health Canada Safety Code 20A and some provincial regulations. In certain circumstances, a software error can lead to a situation where the five minute fluoroscopy audible signal does not sound, as required in Health Canada Safety Code 20A and some provincial regulations. No injuries attributed to the problem are reported.