Health Products

MICROFUGE 20R 120 V 60 HZ WITH FA241.5P IVD (2016-03-08)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Hospitals, Healthcare Professionals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
MRB15K072
Manufacturer
Beckman Coulter
Inc.
250 S. Kraemer Blvd.
92821
UNITED STATES
A. MICROFUGE 20R 120 V 60 HZ WITH FA241.5P IVDB30147

Introduction

Affected products 

A. MICROFUGE 20R 120 V 60 HZ WITH FA241.5P IVD

Reason

Beckman Coulter has identified that the power boards on the affected units may have an incorrect component. As a result, the power board may short circuit resulting in an error code 22, which will cause the centrifuge to become inoperable.