INTELLISPACE PORTAL (2016-03-09)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: Healthcare Professionals, General Public, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
All lotsA. INTELLISPACE PORTAL
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684
NETHERLANDS
881001

Introduction

Affected products

A. INTELLISPACE PORTAL

Reason

The AutoSPECT Pro application was only designed to reconstruct cardiac SPECT data obtained with detectors positioned at 90° or 180°. Relative to one another. However, certain Gamma cameras, e.g., the Marconi Axis and Irix cameras, allow for other relative detector angles. Philips has become aware that data acquired at these other angles will not be correctly reconstructed by AutoSPECT Pro, and the results will likely be erroneous.