Health Products

ADVIA CHEMISTRY - TRIGLYCERIDES CONCENTRATED REAGENTS (2016-03-09)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
348297
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES
B. ADVIA CHEMISTRY - TRIGLYCERIDES_2 CONCENTRATED REAGENTS (TRIG_C)10697575
348297
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES
A. ADVIA 1200 CHEMISTRY - TRIGLYCERIDES_2 CONCENTRATED REAGENTS (TRIG_C)10697575

Introduction

Affected products 
  1. ADVIA 1200 CHEMISTRY - TRIGLYCERIDES_2 CONCENTRATED REAGENTS (TRIG_C)
  2. ADVIA CHEMISTRY - TRIGLYCERIDES_2 CONCENTRATED REAGENTS (TRIG_C)

Reason

Siemens Healthcare Diagnostics has confirmed the ADVIA Chemistry Triglyceride concentrated reagent (trig_c) does not meet instructions for use (IFU) linearity claim at the upper limit of the assay range as it approaches end of shelf life for lots 348297 and 359932.

Internal testing demonstrated that a patient bias (-12% to -15%) may be seen at concentration levels between 450 to 550mg/dl (5.085 - 6.215mmol/l).  Linearity is reduced by approximately 30% at higher triglyceride concentrations up to 1200mg/dl (13.56mmol/L).

Quality control (QC) material may not detect this issue depending on the concentrations used during QC testing.