Health Products

APOLLO AND JUNO DRF (2016-02-17)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
AllB. JUNO DRF
Manufacturer
Villa Sistemi Medicali S.P.A.
via Delle Azalee 3
Buccinasco
20090
ITALY
9784729831
S/N 08060167APOLLO-MAIN UNIT
Manufacturer
Villa Sistemi Medicali S.P.A.
via Delle Azalee 3
Buccinasco
20090
ITALY
9784AXBYVT

Introduction

Affected products
  1. APOLLO-MAIN UNIT
  2. JUNO DRF

Reason

An operator got hurt while relocating a patient from table onto the patient's bed. Standing behind the table top between the left table top support arm and the detector holder, he got squeezed between the left table top support arm and the detector holder.