The current recommendation of spraying EVICEL (Fibrin Sealant) with the 45 cm tip during laparoscopy surgery is at a pressure range of 1.4 to 1.7bar. The European Medicines Agency have clarified their recommendation for pressure to be used when spraying EVICEL with the 45 cm tip during laparoscopic surgery from a range of 1.4-1.7 bar to 1.4 bar only. A similar approach has been taken in Canada. Thus, the labelling of the pressure regulators will be updated and the sticker label on each pressure regulator will be replaced to reflect this change.