CDI BLOOD PARAMETER MONITORING SYSTEM 500 (2016-03-16)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
3030
Manufacturer
Terumo Cardiovascular Systems Corp.
6200 Jackson Road
Ann Arbor
Michigan
UNITED STATES
CDI BLOOD PARAMETER MONITORING SYSTEM 500500AHCT

Introduction

Affected Products

CDI BLOOD PARAMETER MONITORING SYSTEM 500

Reason

Specific CDI Blood Parameter Monitoring System 500 devices are being voluntarily recalled because the BPM sensor head assembly's thermistor, which provides the blood temperature value that results in accurate display values on the monitor, does not meet specification. This may cause inaccurate temperature measurement and inaccurate analyte display values on the CDI System 500 monitor.
Terumo Cardiovascular Systems (Ryan Medical Distributors Inc.) Received complaints of inaccurate temperature measurements for specific devices distributed since November 5, 2015. These include both new production devices and recently serviced devices.