ADVIA 560 HEMATOLOGY ANALYZER (2016-03-11)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
S020099
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES
ADVIA 560 HEMATOLOGY ANALYZER11170842

Introduction

Affected products 
  1. ADVIA 560 HEMATOLOGY ANALYZER

Reason

Siemens has identified that software version 1.4.2133 does not trigger the following flags:
- the W warning flag for detecting small sample volumes when using the international software configuration.
If there is not enough sample volume of the patient's blood when the sample is run, results may be lower than expected. This can occur if the required sample volume as defined in the ADVIA 560 Hematology System operator's guide is not followed.
- the G or L morphology flags for immature granulocytes (IG) and atypical lymphocytes (AtLP), respectively.
The results from patient samples which have immature granulocytes or atypical lymphocytes will not generate the flags when they should. As defined in the ADVIA 560 Hematology System operator's guide, the immature granulocytes and atypical lymphocyte morphology flags should be raised when the presence of morphologically abnormal cells is suspected indicating that a manual count on a stained smear needs to be performed.