Health Products

AISYS CS2 ANESTHESIA MACHINE MAIN UNIT (2016-03-10)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: Hospitals, General Public, Healthcare Professionals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturer.AISYS CS2 ANESTHESIA MACHINE MAIN UNIT
Manufacturer
Datex-Ohmeda
3030 Ohmeda Drive, P.O. Box 7550,
Madison
53707-7550
Wisconsin
UNITED STATES
1011-9050-000

Introduction

Affected Products

AISYS CS2 ANESTHESIA MACHINE MAIN UNIT

Reason

The vacuum suction tubing could be kinked within the machine resulting in suction less than the 20 LPM required by ISO 10079-3. If this issue is left unresolved, it could result in inadequate suction performance to clear the airway and inadequate ventilation of the patient. There have been no injuries reported as a result of this issue.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime