1004127190VIDAS LYME IGG II
Manufacturer
Biomerieux Sa
Chemin De L'orme
Marcy-l'etoile
69280
FRANCE
417401

Introduction

Affected Products

VIDAS LYME IGG II

Reason

Some customers reported that they could not perform calibration. Results of investigation reveal that calibration cannot be performed due to S1 RFV signal which triggers "invalid calibration" error.