Health Products

VIDAS SYSTEM - TOXO IGM (TXM) ASSAY (2016-02-08)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturer.A. VIDAS SYSTEM - TOXO IGM (TXM) ASSAY
Manufacturer
Biomerieux Inc.
595 Anglum Road
Hazelwood
63042
Missouri
UNITED STATES
30202

Introduction

Affected products

A. VIDAS SYSTEM - TOXO IGM (TXM) ASSAY

Reason

An external study showed that performing tests with inactivated sera provides discrepant results. The index results obtained testing inactivated serum are lower compared to values obtained with fresh serum. This could lead for borderline sample to a drift in the overall result interpretation (i.e. from positive to equivocal, from equivocal to negative, from positive to negative).