enGen Laboratory Automation System (2016-03-18)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
J16XQH4J
Manufacturer
Thermo Fisher Scientific Oy
P.O. Box 100, Ratastie 2
Vantaa
01621
FINLAND
A. enGen Laboratory Automation SystemENGEN

Introduction

Affected products 

A. enGen Laboratory Automation System

Reason

Ortho Clinical Diagnostics (Ortho) has initiated this recall pertaining to the enGen Laboratory Automation System configured with Thermo Scientific Recapper Module due to the possibility that the electrical cable that supplies power to the cap Vibratory Feeder of the Recapper Module may loosen from its correct position above the Vibration Feeder support plate and migrate to under the support plate. If this occurs, the uncapped sample tubes may hit the cable, and sample fluid (e.g., serum) may splash from the tube and into another tube or onto the under carriage of the support plate.