Introduction

Affected products

A. SOMATOM Force

Reason

Siemens is informing customers of a potential malfunction when using the scan protocol "neonate head" with the pediatric kernel Hp38. This model is especially used for head examinations and could result in misdiagnosis due to image artifacts.