Health Products

CF InPlex ASR Card (2016-03-31)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
U34AH
Manufacturer
Gen-Probe Incorporated, a wholly owned subsidiary of Hologic, Inc.,
10210 Genetic Center Drive,
San Diego.
92121
California
UNITED STATES
A. CF InPlex ASR Card95-0501

Introduction

Affected products

A. CF InPlex ASR Card

Reason

This voluntary recall is being initiated as a result of complaints of false positive het mutations (9 complaints) and leaking cards (2 complaints) reported to Hologic.  Internal investigations have determined that the false positives and leakage were most likely caused by manufacturing defect at a key component supplier.