UniCel DxH Analysis Systems (2016-04-05)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots
Manufacturer
Beckman Coulter
Inc.
250 S. Kraemer Blvd.
92821
UNITED STATES
B. UNICEL DXH SLIDEMAKER STAINER COULTER CELLULAR ANALYSIS SYSTEM775222
All lots
Manufacturer
Beckman Coulter
Inc.
250 S. Kraemer Blvd.
92821
UNITED STATES
C. UNICEL DXH 600 COULTER CELLULAR ANALYSIS SYSTEMB23858
All lots
Manufacturer
Beckman Coulter
Inc.
250 S. Kraemer Blvd.
92821
UNITED STATES
A. UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM629029

Introduction

Affected products 

A. UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM

B. UNICEL DXH SLIDEMAKER STAINER COULTER CELLULAR ANALYSIS SYSTEM

C. UNICEL DXH 600 COULTER CELLULAR ANALYSIS SYSTEM

Reason

The DxH 800, DxH 600, and DxH Slidemaker Stainer may fail to notify the user of warnings or error conditions for low or depleted reagent conditions, daily checks failures, LIS disconnection, and printer failures at:
- the alert status icon(s) on the system manager
- the audible alarm at the system manager
- the instrument beacon on the specimen processing module or Slidemaker Stainer
the failure may occur when the system manager application is restarted or powered off/on.