Health Products

Cook Inc. catheters, catheter sets and accessories (2016-04-19)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots.
Manufacturer
Cook, Inc.
750 Daniels Way
Bloomington
47404
Indiana
UNITED STATES
H. BEACON TIP ROYAL FLUSH PLUS HIGH-FLOW CATHETERHNR5.0-XXX
All lots.
Manufacturer
Cook, Inc.
750 Daniels Way
Bloomington
47404
Indiana
UNITED STATES
I. SHUTTLE SELECT SLIP-CATH AND SLIP-CATH BEACON TIP CATHETERSCBR5.0-XXX
All lots.
Manufacturer
Cook, Inc.
750 Daniels Way
Bloomington
47404
Indiana
UNITED STATES
F. BEACON TIP WHITE VESSEL SIZING CATHETERNR5.0-35-100-P-10S-PIG-WSC
All lots.
Manufacturer
Cook, Inc.
750 Daniels Way
Bloomington
47404
Indiana
UNITED STATES
O. TRANSLUMINAL BILIARY BIOPSY FORCEPS SETBBFS-100
All lots.
Manufacturer
Cook, Inc.
750 Daniels Way
Bloomington
47404
Indiana
UNITED STATES
G. BEACON TIP TORCON NB ADVANTAGE CATHETERHNBR6.0-XXX
All lots.
Manufacturer
Cook, Inc.
750 Daniels Way
Bloomington
47404
Indiana
UNITED STATES
P. BEACON TIP TORCON NB ADVANTAGE CATHETERHNBR5.0-XXX
All lots.
Manufacturer
Cook, Inc.
750 Daniels Way
Bloomington
47404
Indiana
UNITED STATES
D. SELECTIVE SALPINGOGRAPHY CATHETER WITH BEACON TIPJ-SSG-504000
All lots.
Manufacturer
Cook, Inc.
750 Daniels Way
Bloomington
47404
Indiana
UNITED STATES
E. WHITE LUMAX GUIDING COAXIAL CATHETERLMGRF-7.0C-80-MPA-PULM
All lots.
Manufacturer
Cook, Inc.
750 Daniels Way
Bloomington
47404
Indiana
UNITED STATES
B. LIVER ACCESS AND BIOPSY NEEDLE SETLABS
All lots.
Manufacturer
Cook, Inc.
750 Daniels Way
Bloomington
47404
Indiana
UNITED STATES
K. BEACON TIP CENTIMETER SIZING CATHETERNR5.0-35-100-P-10S-0-CSC-20
All lots.
Manufacturer
Cook, Inc.
750 Daniels Way
Bloomington
47404
Indiana
UNITED STATES
L. HASKAL TRANSJUGULAR INTRAHEPATIC PORTAL ACCESS SETHTPS-XXX
All lots.
Manufacturer
Cook, Inc.
750 Daniels Way
Bloomington
47404
Indiana
UNITED STATES
C. KUMPE ACCESS CATHETER023565-BT
All lots.
Manufacturer
Cook, Inc.
750 Daniels Way
Bloomington
47404
Indiana
UNITED STATES
M. FLUOROSET RADIOGRAPHIC TUBAL ASSESSMENT SETJ-RTAS-100
All lots.
Manufacturer
Cook, Inc.
750 Daniels Way
Bloomington
47404
Indiana
UNITED STATES
N. APRIMA ACCESS NONVASCULAR INTRODUCER SETNPAS-100-HC-NT-U-SST
All lots.
Manufacturer
Cook, Inc.
750 Daniels Way
Bloomington
47404
Indiana
UNITED STATES
A. NEFF D'AGOSTINO PERCUTANEOUS ACCESS SETNPAS-100-D'AGOSTINO-B-050393
All lots.
Manufacturer
Cook, Inc.
750 Daniels Way
Bloomington
47404
Indiana
UNITED STATES
J. BEACON TIP VESSEL SIZING CATHETERNR5.0-35-70-P-10S-PIG-CAVA

Introduction

Affected Products
  • A. NEFF D'AGOSTINO PERCUTANEOUS ACCESS SET
  • B. LIVER ACCESS AND BIOPSY NEEDLE SET
  • C. KUMPE ACCESS CATHETER
  • D. SELECTIVE SALPINGOGRAPHY CATHETER WITH BEACON TIP
  • E. WHITE LUMAX GUIDING COAXIAL CATHETER
  • F. BEACON TIP WHITE VESSEL SIZING CATHETER
  • G. BEACON TIP TORCON NB ADVANTAGE CATHETER
  • H. BEACON TIP ROYAL FLUSH PLUS HIGH-FLOW CATHETER
  • I. SHUTTLE SELECT SLIP-CATH AND SLIP-CATH BEACON TIP CATHETER
  • J. BEACON TIP VESSEL SIZING CATHETER
  • K. BEACON TIP CENTIMETER SIZING CATHETER
  • L. HASKAL TRANSJUGULAR INTRAHEPATIC PORTAL ACCESS SET
  • M. FLUOROSET RADIOGRAPHIC TUBAL ASSESSMENT SET
  • N. APRIMA ACCESS NONVASCULAR INTRODUCER SET
  • O. TRANSLUMINAL BILIARY BIOPSY FORCEPS SET
  • P. BEACON TIP TORCON NB ADVANTAGE CATHETER

Reason

Cook Medical has identified an increase in reports of polymer degradation of the catheter tip, resulting in tip fracture and/or separation. Their preliminary investigation into this matter has identified that environmental conditions, such as storage temperature, humidity, and the use of vaporized hydrogen peroxide (VHP) for whole-room decontamination within healthcare facilities, may be contributing to the occurrence. There may also be other undetermined contributors to this issue. Investigation is ongoing.

Potential adverse events that may occur as a result of catheter polymer degradation could include loss of device function, separation of a device segment leading to medical intervention, or complications resulting from a separated segment. Such complications include device fragments in the vascular system, genitourinary system, or other soft tissues. Fragments within the vascular system could result in embolization to the heart or lungs, or occluding blood flow to end organs.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime