IMMULITE System Assays (2016-04-20)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
411B. IMMULITE 1000 SYSTEM - IGF-1 ASSAY
Manufacturer
Siemens Healthcare Diagnostics Products Limited
Glyn Rhonwy
Llanberis
LL55 4EL
UNITED KINGDOM
LKGF1
411A. IMMULITE SYSTEM - IGF-1 (INSULIN LIKE GROWTH HORMONE)
Manufacturer
Siemens Healthcare Diagnostics Products Limited
Glyn Rhonwy
Llanberis
LL55 4EL
UNITED KINGDOM
LKGF1

Introduction

Affected Products

A. IMMULITE SYSTEM - IGF-1 (INSULIN LIKE GROWTH HORMONE)
B. IMMULITE 1000 SYSTEM - IGF-1 ASSAY

Reason

Manufacturer has determined that it takes 24 minutes of incubation prior to process the samples on the IMMULITE/IMMULITE 1000 System for the patient sample and the pretreatment solution to reach full equilibration when using pretreatment solution (LGFA) lot 055 contained in IGF-I Kit lot 411. If patient samples are run before reaching full equilibration an under-recovery of up to -36% may occur. If the sample is left to incubate for more than 24 minutes prior to testing, no under recovery is observed.