DISCOVERY IGS 730/740 SYSTEM (2016-04-18)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: Healthcare Professionals, General Public, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
0000HXED001317
Manufacturer
GE Medical Systems SCS
283 Rue De La Miniere
Buc
B. DISCOVERY IGS 7405418813
0000HXE9001292
Manufacturer
GE Medical Systems SCS
283 Rue De La Miniere
Buc
DISCOVERY IGS 730 SYSTEM5418812

Introduction

Affected products 
  1. DISCOVERY IGS 730 SYSTEM
  2. DISCOVERY IGS 740

Reason

The following issue is being addressed by GE Healthcare file No. FMI 12246: GE Healthcare has recently become aware of a potential safety issue related to a non-recoverable loss of image acquisition. The affected discovery systems may experience multiple x-ray abort errors during a real-time interventional procedure. This issue could potentially happen before or during a fluoroscopic interventional procedure. There have been no injuries reported as a result of this issue.