REVOLUTION CT - MAIN UNIT (2016-04-21)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Hospitals, Healthcare Professionals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
443342CN1
Manufacturer
GE Medical Systems LLC
3200 North Grandview Blvd
Waukesha
53188
Wisconsin
UNITED STATES
REVOLUTION CT - MAIN UNIT5590000

Introduction

Affected products 
  1. REVOLUTION CT - MAIN UNIT

Reason

GE Healthcare has become aware that an out of tolerance torque tool was used on three joints on revolution CT system during the manufacturing process. If the x-ray tube becomes loose on the CT gantry during operation, this could result in beam tracking errors, potential scan aborts and/or a possible image artifact. There is a secondary safety mechanism to prevent components from being expelled from the gantry if the bolts fail. There have been no injuries reported as a result of this issue.