Health Products

IMAGE SUITE 3.0/4.0 SOFTWARE (2016-04-20)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
52412542B. IMAGE SUITE 3.0 WITH MINI-PACS OPTIONS AND HOT FIX 3004 SOFTWARE PATCH
Manufacturer
Carestream Health
150 Verona Street
Rochester
14608
New York
UNITED STATES
1036490
52412542C.  IMAGE SUITE 3.0 WITH MINI-PACS OPTIONS AND HOT FIX 3004 SOFTWARE
Manufacturer
Carestream Health
150 Verona Street
Rochester
14608
New York
UNITED STATES
1036508
85013463IMAGE SUITE 4.0 WITH MINI-PACS AND DICOM STORE
Manufacturer
Carestream Health
150 Verona Street
Rochester
14608
New York
UNITED STATES
1056191

Introduction

Affected products
  1. IMAGE SUITE 4.0 WITH MINI-PACS AND DICOM STORE
  2. IMAGE SUITE 3.0 WITH MINI-PACS OPTIONS AND HOT FIX 3004 SOFTWARE PATCH
  3. IMAGE SUITE 3.0 WITH MINI-PACS OPTIONS AND HOT FIX 3004 SOFTWARE

Reason

Carestream has identified a software issue that if the image suite software receives Computed Tomography (CT) and Magnetic Resonance (MR) images with the image orientation (patient) tag specified, and the overlay is configured to display this tag, the L (left) and R (right) and A (anterior) and P (posterior) overlay text could be displayed reversed.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime