B. IMAGE SUITE 3.0 WITH MINI-PACS OPTIONS AND HOT FIX 3004 SOFTWARE PATCH
Manufacturer
Carestream Health
150 Verona Street
Rochester
14608
New York
UNITED STATES
1036490
52412542
C. IMAGE SUITE 3.0 WITH MINI-PACS OPTIONS AND HOT FIX 3004 SOFTWARE
Manufacturer
Carestream Health
150 Verona Street
Rochester
14608
New York
UNITED STATES
1036508
85013463
IMAGE SUITE 4.0 WITH MINI-PACS AND DICOM STORE
Manufacturer
Carestream Health
150 Verona Street
Rochester
14608
New York
UNITED STATES
1056191
Introduction
Affected products
IMAGE SUITE 4.0 WITH MINI-PACS AND DICOM STORE
IMAGE SUITE 3.0 WITH MINI-PACS OPTIONS AND HOT FIX 3004 SOFTWARE PATCH
IMAGE SUITE 3.0 WITH MINI-PACS OPTIONS AND HOT FIX 3004 SOFTWARE
Reason
Carestream has identified a software issue that if the image suite software receives Computed Tomography (CT) and Magnetic Resonance (MR) images with the image orientation (patient) tag specified, and the overlay is configured to display this tag, the L (left) and R (right) and A (anterior) and P (posterior) overlay text could be displayed reversed.