Health Products

ARTHROWAND, TOPAZ MICRODEBRIDER W/INTEGRATED CABLE W/O SUCTION (2016-03-04)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
1032755ARTHROWAND, TOPAZ MICRODEBRIDER W/INTEGRATED CABLE W/O SUCTION
Manufacturer
Arthrocare Corporation,
7000 W. William Cannon Drive,
Austin
78735
Texas
UNITED STATES
AC4040-01

Introduction

Affected products 
  1. ARTHROWAND, TOPAZ MICRODEBRIDER W/INTEGRATED CABLE W/O SUCTION

Reason

During functional testing, it was determined that product failed to meet shelf-life at the expiration date indicated on the product labels. This action is due to product misbranding. The expiration date is not supported by the real time aging collected. There are no associated immediate or long range health consequences anticipated from the use of or exposure to the defective device.