Health Products

COBAS H 232 INSTRUMENTS (2016-04-26)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Hospitals, Healthcare Professionals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 100 numbers, contact manufacturer. 
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY
COBAS H 232 INSTRUMENTS4901142190

Introduction

Affected products 
  1. COBAS H 232 INSTRUMENTS

Reason

Contamination of the mirror from the optical system can lead to false positive signals in the early phase of measurement which can lead to the following failure pattern on a cobas h 232: "TnT elevated" displayed during the measurement but the final results is trop T (less than) 50ng/l, IQC level-II fails, E-402 during a measurement.