The system may sporadically apply the default x-ray exposure parameters for an adult (patient type: "normal"), even though the patient type "newborn" was selected and is displayed in the generator control area of the Eleva user interface.
Under certain conditions, the detector might not be ready for examination. Released x-ray might lead to an image with artifacts and a retake is required. While the attachment process is running, the detector might be too short in front of the IR (infrared) sensor and the problem of the washed out images can appear if an exposure is taken right after that.