MOBILEDIAGNOST WDR (2016-04-13)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots 
Manufacturer
Philips Medical Systems Dmc Gmbh
Rontgenstrabe 24
Hamburg
22335
GERMANY
A. MOBILEDIAGNOST WDR9890 010 8952X

Introduction

Affected products 
  1. MOBILEDIAGNOST WDR

Reason

Philips has become aware of the following issues:
  1. The system may sporadically apply the default x-ray exposure parameters for an adult (patient type: "normal"), even though the patient type "newborn" was selected and is displayed in the generator control area of the Eleva user interface.
  2. Under certain conditions, the detector might not be ready for examination. Released x-ray might lead to an image with artifacts and a retake is required. While the attachment process is running, the detector might be too short in front of the IR (infrared) sensor and the problem of the washed out images can appear if an exposure is taken right after that.