Health Products

KT-DTM LAMP (2016-04-22)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturer.A. KT-DTM LAMP
Manufacturer
KLOX TECHNOLOGIES INC.
275 ARMAND FRAPPIER
LAVAL
H7V 4A7
Quebec
CANADA
KLLA0305-01

Introduction

Affected products
  1. KT-DTM LAMP

Reason

This information is initiated as a result of an investigation following the development of a new KT-D design in which the power density of the new design did not yield the same results made by two different testing facilities and with controlled and transferred methods. A detailed investigation has confirmed that there were issues related to the power density method, its transfer and its validation. It is possible that the power density of the KT-D lamp may have a higher value than the value currently approved in production established at 55.6 to 71.6 mw/cm2