Health Products

Lifecodes HLA-C eRES SSO Typing Kit (2016-05-02)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
Lot # 10235BA.
Manufacturer
Immucor Transplant Diagnostics, Inc.
550 West Avenue
Stamford
06902
Connecticut
UNITED STATES
628921

Introduction

Affected products 
  1. Lifecodes HLA-C eRES SSO Typing Kit

Reason

It has been determined that the threshold set for HLA-C eRES probes 380, 382, 384 and 385 are set high and may result in false negative probe results. The probes resolve the following CWD alleles probe 380 resolves c*03:06, probe 382 resolves c*12:13, probe 384 resolves c*01:08 and probe 385 resolves c*08:06.