UNICEL DXH 600/800 COULTER CELLULAR ANALYSIS SYSTEM (2016-05-02)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
AZ02609
Manufacturer
Beckman Coulter
Inc.
250 S. Kraemer Blvd.
92821
UNITED STATES
B. UNICEL DXH 600 COULTER CELLULAR ANALYSIS SYSTEMB23858
AU38544
Manufacturer
Beckman Coulter
Inc.
250 S. Kraemer Blvd.
92821
UNITED STATES
A. UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM 629029

Introduction

Affected products
  1. UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM

  2. UNICEL DXH 600 COULTER CELLULAR ANALYSIS SYSTEM

Reason

Beckman Coulter has confirmed an issue where the Red Blood Cell (RBC) aperture baths can contain residue in the sweep flow fittings. This residue may potentially create increased background failures, flagged results and vote outs for RBC and platelet (plt) parameters.