SONALLEVE MR-HIFU FIBROID THERAPY (2016-04-28)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
6102SONALLEVE MR-HIFU FIBROID THERAPY
Manufacturer
PHILIPS OY - PHILIPS MEDICAL SYSTEMS MR FINLAND
AYRITIE 4
VANTAA
01510
FINLAND
994996

Introduction

Affected Products

SONALLEVE MR-HIFU FIBROID THERAPY

Reason

When performing sonication with the Sonalleve MR-HIFU fibroid therapy system (with r2.1 l3 application software) with treatment cells of 2mm in size, there is a possibility that the depth of sonication is not shown on the user interface due to a software defect. A maximum positioning error of 4cm in longitudinal direction is possible. This may occur either in the uterine fibroid or the bone application.