Health Products

OLERUP SSP HLA-A (2016-05-04)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
35X
Manufacturer
Olerup SSP AB
Franzengatan 5
Stockholm, Sweden, 11251
A. OLERUP SSP HLA-A101.411-06U

Introduction

Affected products
  1. OLERUP SSP HLA-A

Reason

In kits of A*01, lots 41V, 35X and 83Y, the A*01:04N allele is not amplified by primer-mix 5 as stated in product documentation. The false negative amplification for the A*01 kit yields a revised reactivity pattern for the A*01:04n allele that is identical to A*01:01 alleles.