Health Products

FT4 II (FREE THYROXINE II) (2016-05-11)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
188609A. FT4 II (FREE THYROXINE II)
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY
6437281190

Introduction

Affected products
  1. FT4 II (FREE THYROXINE II)

Reason

In the software screens on cobas e 602, the expiry date of FT4 II reagent kit lot 188609 could be displayed 11/2016, instead of the correct date 10/2016 which is the expiry date of the box and the reagent kit label. When this occurs, the false expiry date is used with the reagent kits.