Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY
A. TRIGL (TRIGLYCERIDES) FOR COBAS C701 ANALYZER5171407190

Introduction

Affected products 
  1. TRIGL (TRIGLYCERIDES) FOR COBAS C701 ANALYZER

Reason

The manufacturer has identified during an internal investigation on a unique customer complaint, that a change in the Cobas C701-702 application prozone limit check was required for the triglycerides assay to address a potential for low results without the expected flag. The application package has been modified and has been provided accordingly to add the new prozone limit check.