The manufacturer has identified during an internal investigation on a unique customer complaint, that a change in the Cobas C701-702 application prozone limit check was required for the triglycerides assay to address a potential for low results without the expected flag. The application package has been modified and has been provided accordingly to add the new prozone limit check.