PHADIA 100/250 SYSTEM (2016-04-21) Apr 21, 2016 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type III
Lot or serial number Companies Brand name Model or catalog number Not Applicable Manufacturer PHADIA AB RAPSGATAN 7 UPPSALA SWEDEN B. PHADIA 250 SYSTEM - INSTRUMENT REAGENTS 83-1017-01 Not Applicable Manufacturer PHADIA AB RAPSGATAN 7 UPPSALA SWEDEN C. PHADIA 250 SYSTEM - ELIA CCP ASSAY 14-5515-01 Not Applicable Manufacturer PHADIA AB RAPSGATAN 7 UPPSALA SWEDEN A. PHADIA 100 SYSTEM - ELIA CCP 14-5515-01
Affected products
PHADIA 100 SYSTEM - ELIA CCP
PHADIA 250 SYSTEM - INSTRUMENT REAGENTS
PHADIA 250 SYSTEM - ELIA CCP ASSAY
When using EliA CCP well in combination with EliA IgG conjugate, anti-CCP signals can be elevated. The positive bias will change a small percentage of test results from negative to equivocal and equivocal or high negative to low positive.