Health Products

Dialog+ Dialysis Machines (2016-03-18)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
B. Braun Avitum Ag
Schwarzenberger Weg 73-79
Melsungen
34212
GERMANY
A. Dialog+ Dialysis Machines710200K

Introduction

Affected products 

A. Dialog+ Dialysis Machines

Reason

Flanges assembled into conductivity cells show hairline cracks.  If a crack occurs it may lead to a leakage and result in a balance deviation.  Affected machines will have improper filtration rate.