B. MAGNA-FX CANNULATED SCREW FIXATION SYSTEM AND INSTRUMENTATION
00114205124
63225662
Manufacturer
Zimmer Inc.
1800 West Center Street
Warsaw
46580
UNITED STATES
C. TRILOGY ACETABULAR SYSTEM-BONE SCREWS
6250-45-25
63290189
Manufacturer
Zimmer Inc.
1800 West Center Street
Warsaw
46580
UNITED STATES
A. COMPRESSION SCREW
47116202400
Introduction
Affected products
A. COMPRESSION SCREW
B. MAGNA-FX CANNULATED SCREW FIXATION SYSTEM AND INSTRUMENTATION
C. TRILOGY ACETABULAR SYSTEM-BONE SCREWS
D. COMPRESSION SCREW
Reason
These screws are packaged with a double barrier system, two plastic trays and two Tyvek lids, to provide sterile integrity. Process monitoring conducted as part of the standard packaging process identified that in some instances, a small hole may be present in the inner tray. The holes were found to be caused by a single tool used by the tray supplier in the manufacturing process and estimated to be present in less than 10% of the affected devices. The outer tray is not affected and the device's sterile integrity remains until the outer tray is opened. No product complaints have been reported for this issue.