OPTIMA MR430S 1.5T - MAIN UNIT (2016-05-13)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: Hospitals, Healthcare Professionals, General Public
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
0004212012-001
Manufacturer
GE Medical Systems, LLC
301 BALLARDVALE ST., SUITE 4
WILMINGTON
01887
Massachusetts
UNITED STATES
A. OPTIMA MR430S 1.5T - MAIN UNITOPTIMA MR430S 1.5T (5000-0002)

Introduction

Affected products

A. OPTIMA MR430S 1.5T - MAIN UNIT

Reason

GE Healthcare has recently become aware of a potential safety issue with gas venting. A site reported a magnet quenching with subsequent venting of the cryogen gas into the magnet room. Cryogen gas that vents into the magnet room could cause critical deprivation of oxygen to users and patients. There have been no injuries reported as a result of this issue.
At this site, a magnet heater probe connection became disconnected. This resulted in ice build-up within the magnet blocking the cryogen vent. Since the vent was blocked by ice, the later magnet quench caused gas to be vented into the magnet room.